Operation

Guidelines for operation

C A U T I O N

Personal injury hazard. The glass bulb or shank on the sensor can break. Handle the sensor carefully to prevent
injury.

• Remove the protective cap before the sensor is put into the process.
• When the sensor is removed from the process for >1 hour, fill the protective cap with pH 4 buffer

(recommended) or tap water and put the cap on the sensor. Repeat every 2–4 weeks for extended
storage.

User navigation

Refer to the User interface and navigation section of the controller documentation for keypad
description and navigation information.

Configure the sensor

Use the CONFIGURE menu to enter identification information for the sensor and to change options
for data handling and storage.

1. Push the menu key and select SENSOR SETUP>[Select Sensor]>CONFIGURE.
2. Use the arrow keys to select an option and push enter. To enter numbers, characters or

punctuation, push and hold the up or down arrow keys. Push the right arrow key to advance to
the next space.

Option

Description

EDIT NAME

Changes the name that corresponds to the sensor on the top of the measure screen.
The name is limited to 10 characters in any combination of letters, numbers, spaces
or punctuation.

SENSOR S/N

Allows the user to enter the serial number of the sensor, limited to 16 characters in
any combination of letters, numbers, spaces or punctuation.

DISPLAY FORMAT

For pH sensors only—changes the number of decimal places that are shown on the
measure screen.

TEMP UNITS

Sets the temperature units to °C (default) or °F

TEMP ELEMENT

pH sensors—sets the temperature element to PT100 or PT1000 for automatic
temperature compensation. If no element is used, the type can be set to MANUAL
and a value for temperature compensation can be entered. ORP sensors—
temperature compensation is not used. A temperature element can be connected to
measure temperature.

FILTER

Sets a time constant to increase signal stability. The time constant calculates the
average value during a specified time—0 (no effect) to 60 seconds (average of signal
value for 60 seconds). The filter increases the time for the sensor signal to respond to
actual changes in the process.