User manual, Specifications, Ph/‍orp sensors – Hach-Lange POLYMETRON 9500 Ultrapure pH_ORP Module User Manual User Manual

Page 3: Antimony electrode sensors, English, Ph/orp sensors

Advertising
background image

Specifications

Specifications are subject to change without notice.

pH/ORP sensors

8350.3

8350.4

8350.5

8351

Application

Measuring pH in

samples containing

hydrofluoric acid

Measuring pH at

high temperatures

Measuring pH in a

waste water

environment

Measuring ORP

Material

PPS

PPS

CPVC

PPS

Measurement range

0—12 pH

0—14 pH

0—12 pH

± 1500 mV

Maximum
temperature

110 °C

110 °C

80 °C

110 °C

Maximum pressure

10 bar

10 bar

10 bar

10 bar

Repeatability (week)

< 0.05 pH

< 0.05 pH

< 0.05 pH

< 0.05 pH

Offset @ pH7

± 0.34 pH

± 0.34 pH

± 0.34 pH

N/A

Slope

56—61 mV/pH

56—61 mV/pH

56—61 mV/pH

N/A

Reference
impedance @ 25 °C

< 50 Kohm

< 50 Kohm

< 50 Kohm

< 50 Kohm

Glass impedance @
25 °C

100—150 Mohm

150—500 Mohm

50—250 Mohm

N/A

8362

8417

Application

Measuring pH in pure or ultra pure

water

Measuring pH in industrial and

waste water treatment plants

Material

316L stainless steel

Glass membrane, ceramic junction

Measurement range

2—12 pH

0—14 pH

Maximum temperature

80 °C

110 °C

Maximum pressure

16 bar @ 25 °C;

6 bar @ 100 °C

10 bar @ 25 °C

Repeatability (24 hours)

< 0.01 pH

0.02 pH

Input impedance

> 10

6

Mohm

100 Mohm

Antimony electrode sensors

8346 / 8347

Application

Measuring pH or ORP in dirty water

Material

Polypropylene

Maximum temperature

80 °C (pH measurements with 8346)

120 °C (pH measurements with 8347)

50 °C (ORP measurements)

Sample pressure

Atmospheric

Maximum pressure (electrolyte container)

5 bar

English 3

Advertising