Ma outputs, Output parameters, Event indication – Hach-Lange POLYMETRON 9245 Basic User Manual User Manual

Page 16: Section entitled

Advertising
background image

Alarm

Yes

Activate the warning alarm

No

Deactivate the warning alarm

Accept

Manual

When the alarm is triggered, turn it off by pressing the Enter function key

Auto

When the alarm is triggered, it will turn itself off only when the reason for the
alarm being triggered is no longer valid

Relay

N.O.

Normally open

N.C.

Normally closed

5. System alarm:

Alarm

Yes

Activate the system alarm

No

Deactivate the system alarm

Relay

N.O.

Normally open

N.C.

Normally closed

mA outputs

1. Select the mA OUTPUTS option to set up the parameters for all the analog outputs.
2. From the list available, select the mA output you wish to set.

Output parameters

1. Analog output parameters:

Attribute

Choose the attribute that triggers the analog output.

Note: The attribute variable cannot be applied to Output 0, which is reserved for the continuous live output signal.

Type

Choose either a 0-20 mA or 4-20 mA analog output

Mode

Select the mode. This parameter is only selectable if the attribute is set to a measurement channel

Low

The value corresponding to the low end of the scale

Middle

The value corresponding to the mid-point of the scale (only available in dual mode)

High

The value corresponding to the high end of the scale

Event indication

1. Select the EVENT INDICATION option in the mA Outputs menu to display the options available

for event setting.

2. Select the option for which you want to set an event.
3. Define the attribute for the event. This is one of the outputs or None.
4. The MODE can be either a Preset val or Frozen.
5. If a preset value is chosen, you will be required to enter the value of the analog output that will be

forced, when that event occurs.