Overview
Figure 1 Front and rear panels
1 Sodium ion selective electrode
8 Reagent shelf
2 Reference electrode
9 Frame for panel mounting
3 Sample flow adjustment valve
10 Reagent canister holder
4 User interface
11 Local controller box
5 Overflow vessel
12 Electrolyte reservoir
6 Measuring cell
13 Pump box
7 Door lock
14 Sample inlet valve
Installation
W A R N I N G
The analyzer should only be assembled by qualified staff. Mains power should only be connected once
installation has been completed and checked
Mounting the analyzer
C A U T I O N
Wherever the analyzer is to be mounted, it is important to note that it must be placed in an upright position with
the transmitter at the top. It is recommended to use a spirit level to ensure that the analyzer is correctly positioned
and not leaning to one side or forward. This is essential to guarantee the accuracy of the analyzer.
Panel mounting
Mount the analyzer using the fixation holes located around the outside of the analyzer.
Wall mounting
Use the wall mounting kit to fix the instrument to the wall. The distance between the two pieces is
460 mm.