Basic user manual, Specifications, English – Hach-Lange POLYMETRON 9245 Basic User Manual User Manual

Page 3

Advertising
background image

Specifications

Specifications are subject to change without notice.

Specification

Details

Measuring range

0 to 10,000 ppb freely programmable
0 to 200 ppm with K-Kit option

Accuracy

Non-cationic application: ± 0.1 ppb or ± 5% of reading, whichever is greater
Cationic application: ± 2 ppb or 5% reading, whichever is greater

Repeatability

< 0.02 ppb or 1.5% reading, whichever is greater within 10°C variation

Response time

0.1 to 10 ppb T(90%) = 180 secs

Electrode type

pH glass electrode

Number of channels

1 channel

Interference phosphate
10 ppm

Measurement variation less 0.1 ppb

Sample temperature
interference

< 0.5% / °C

Typical environment

Power station; Indoor; Demineralized water plant or instrumentation room

Suspended solids

< 2 NTU, no oil, no grease

Storage

-20 to 60°C (2 to 140°F)

Relative humidity

10 to 80%

Ambient temperature

5 to 50°C (41 to 122°F)

Sample temperature
variation

Stabilization in 10 mins from 15°C to 30°C

pH range of sample

Non-cationic application: 6 to 10 pH
Cationic application: 2 to 10 pH

Flow rate

6 to 9 L/hour

Pressure

0.2 to 6 bar (3 - 87 psi)

Acidity

Less than 250 ppm (equivalent CaCO

3

)

Power supply voltage
fluctuation

± 10%

Over voltage category

2 (according to standard EN 61010-1)

Pollution degree

2 (according to standard CEI 664)

Altitude

< 2000 m

Measurement category

Cat II, Class 1 (overvoltage < 1500V)

Maximum panel dimension
(H x L x D)

850 x 450 x 252.5mm [33.46 x 17.71 x 9.94in]

Inlet

Simple fittings for 6 mm O.D. tubing or ¼" O.D. in PE-low density
¼" OD in PHED-PTFE-SS as option

Outlet

Barbed stem for 12 mm (½" I.D.) hose

English 3