External communications connection, Input/‍output connections, Sample tubes installation – Hach-Lange POLYMETRON 9245 Basic User Manual User Manual

Page 8: Input/output connections

Advertising
background image

External communications connection

1. Run the communications cable through an external cable gland on

the bottom of the cabinet, and into the analyzer.

2. Pass it through the cable gland located right and farthest from you

on the base of the transmitter, so it appears inside the transmitter
through the left front cable gland.

3. Connect the communication cable as indicated. Connection is the

same on the CPU board for both the JBUS/ MODBUS and
PROFIBUS options.

4. Close the transmitter door and secure in place with the 4 screws.
5. Put the local controller box back in its normal position and secure in

place with the 2 holding screws.

Input/Output connections

Before using any of the cable glands, perforate first with a screwdriver. To ensure a good seal, the
external diameter of the cables should be between 5 and 7 mm. The nomenclature given in the
connections column of the following table refers to the same nomenclature that is printed on the I/O
board against each available connection.

Connections

Function

Re1 to Re2

User relays - see

Alarms

on page 15 for more information on the relays

Re7

Warning alarm

Re8

System alarm

In2

For remote calibration

In7

To put the analyzer on standby

Iout0

Used for the current measurement signals

Iout1 to Iout3

Can be freely linked to different parameters like measurement, temperature - refer to the
section entitled

mA outputs

on page 16 for details

On completion, close the local controller box (No. 11 in

Figure 1

on page 6) and secure in place with

the 6 screws.

Sample tubes installation

Connecting the sample
Use new tubes for connections during installation

• Exterior Ø: 6 mm exactly (or ¼'')
• Material: polyethylene or PTFE or FEP
• Flow rate: 6 to 9 L/hour
• Pressure: 0.2 to 6 bar (8-100 psig)
• Sample acidity: Sample acidity should not be more than 300 ppm CaCO

3

• Temperature: 5 to 45°C

Note: For continuous monitoring of sample with a temperature between 5-10°C it is recommended that the
installation should include a static heat-exchanger system.

At this stage of the installation, make sure that the flow valve is closed. Connect the pipes by
inserting them into the quick release connections found on the bottom of the analyzer under the
sampling block (No. 14 in

Figure 1

on page 6). Be sure that the sample line is correctly flushed

before any connection to avoid particle injection into the hydraulic system. If particulate matter is
present in the sample, pre-filtration is necessary. A filter should be inserted in the sample line. One is
available as an option.

8 English

Advertising