20

Operation

SET NH4–N CONC

Entry and display of the NH

4

–N reference value (lab value)

-or-

CAL.CONFIG.>MATX1 K+

1 point matrix correction with potassium (see

section 4.6.3.2 on page 24

)

DATE

Displays the date of the current correction

CONC MEAS 1

Saving the currently measured measurement

SET K+ CONC

Entry and display of the K+ reference value (lab value)

SET NH4–N CONC

Entry and display of the NH

4

–N reference value (lab value)

-or-

CAL.CONFIG.>MATX2

2 point matrix correction (see

section 4.6.3.3 on page 25

)

CONC MEAS 1

Saving the currently measured measurement

DATE

Displays the date of the current correction

SET NH4–N CON

Entry and display of the NH

4

–N reference value (lab value)

CONC MEAS 2

Saving the second measured measurement

DATE

Displays the date of the current correction

SET NH4–N CONC

Entry and display of the NH

4

–N reference value (lab value)

-or-

CAL.CONFIG.>MATX2 K+

2 point matrix correction with potassium (see

section 4.6.3.4 on page 26

)

Note: Carry out the matrix correction only as a standard addition in the sample matrix.

CONC MEAS 1

Saving the currently measured measurement

DATE

Displays the date of the current correction

SET K+ CONC

Entry and display of the K+ reference value (lab value)

SET NH4–N CONC

Entry and display of the NH

4

–N reference value (lab value)

CONC MEAS 2

Saving the second measured measurement

DATE

Displays the date of the current correction

SET K+ CONC

Entry and display of the K+ reference value (lab value)

SET NH4–N CONC

Entry and display of the NH

4

–N reference value (lab value)

CALIBRATE(continued)

CAL.CONFIG.>

VALUE CORR

Value correction (see

section 4.6.3.5 on page 28

)

Once value correction is complete, the correction data are displayed in the form of
the MATX2

1st NH4–N NH4DSC

Enter the NH

4

–N concentration (NH4D sc) of

the 1st point

1st K+ NH4DSC

Enter the K+ concentration (NH4D sc) of the
1st point

1st NH4–N LABOR

Enter the NH

4

–N concentration (lab) of the 1st

point

2nd NH4–N NH4DSC

Enter the NH

4

–N concentration (NH4D sc) of

the second point

2nd K+ NH4DSC

Enter the K+ concentration (NH4D sc) of the
second point

2nd NH4–N LABOR

Enter the NH

4

–N concentration (lab) of the

second point

-or-

CAL.CONFIG.>
PREVIOUS CAL

Selection of one of the last four matrix and value corrections performed

-or-

4.5 Sensor setup menu (continued)