Section 1 specifications – Hach-Lange NH4D sc Ammonium Sensor User Manual

Page 5

Advertising
background image

5

Section 1

Specifications

These are subject to change without notice.

General Information

Measuring method

Ion-selective electrodes for ammonium and potassium, pHD reference electrode
and temperature sensor

Measuring range

0.2 to 1000 mg/l [NH

4

–N]

Minimum detection limit

0.2 mg/l [NH

4

–N]

Precision

5% of the measured value + 0,2 mg/l

1

Reproducibility

5% of the measured value + 0,2 mg/l

1

Response time (90%)

< 2 min

Measuring interval

Continuous

pH range

pH 5 to pH 9

Calibration methods

Sensor code for sensor cartridge

1 and 2-point matrix correction

Power consumption

1 Watt

Power supply

via sc controller (sc100 and sc1000)

Data transfer

via sc controller (sc100 and sc1000)

Ambient data

Typical environment

Municipal waste water applications

Storage temperature

Sensor: –20 to 60 °C (–4 to 140 °F)

Sensor cartridge: 5 to 40 °C (41 to 104 °F)

Operating temperature

Air: –20 to 45 °C (–4 to 113 °F)

Sample temperature

+2 to 40 °C (35 to 104 °F)

Max. flow velocity

< 4 m/s

Max. sensor immersion
depth/pressure

Can be immersed 0,3 to 3,0 m/max. pressure: 0,3 hPa (4,4 psi).

Max. air pressure for cleaning unit
option

3,10 bar (45 psi)

General sensor information

Sensor dimensions

360,9 mm x 48,3 mm (14,21 x 1,9 in) (L x Ø)

See

Figure 1 on page 6

.

Sensor cable length

Standard: 10 m

Optional extension cables are available in the following lengths:
5, 10, 15, 20, 30, 50 m. Total maximum length: 100 m

Sensor weight

Approx. 1326 g (46,77 oz)

Wetted materials

Only for immersed installation: Sensor body made of stainless steel 316 with
ends made of Ryton PPS

®2

.

Installation angle

45° +/– 15° vertical in flow direction

1

with standard solutions

2

Ryton

®

is a registered trademark of Phillips 66 Co.