Operation

4.6.3 Matrix correction

4.6.3.1 MATX1 correction (one point matrix correction)

Proceed as follows to perform the MATX1:

1.

Select

SENSOR SETUP>NH4D SC>CALIBRATE> CAL CONFIG

.

2.

Select

MATX1

in the selection window and press

ENTER

.

3.

Select

CONC MEAS 1

.

The currently measured ammonium and potassium values are
displayed. The drift indicates whether the measured value is stable.

4.

Wait until the measured value is stable and confirm by pressing

ENTER

(Drift should be < 0,03 mg/l). The values for ammonium

and potassium are saved.

5.

Immediately after saving, take a water sample for laboratory
analysis from as close to the sensor as possible.

6.

Analyze the sample immediately after taking the sample as its
ammonium content can change quickly.

After determining the laboratory reference value, proceed as
follows:

7.

Select

SENSOR SETUP>NH4D SC>CALIBRATE>

SET NH4–N CONC

.

8.

Enter the laboratory value (reference value) for NH

4

–N and

press

ENTER

to confirm. Confirmation of the entered laboratory

value activates the matrix correction.

4.6.3.2 MATX1 K+ correction

Proceed as follows to perform the MATX1 K+:

1.

Select

SENSOR SETUP>NH4D SC>CALIBRATE> CAL CONFIG

.

2.

Select

MATX1 K+

in the selection window and press

ENTER

.

3.

Select

CONC MEAS 1

.

The currently measured ammonium and potassium values are
displayed. The drift indicates whether the measured value is stable.

4.

Wait until the measured value is stable and confirm by pressing

ENTER

(Drift should be < 0,03 mg/l). The values for ammonium

and potassium are saved.

5.

Immediately after saving, take a water sample for laboratory
analysis from as close to the sensor as possible.

CALIBRATE

CAL. CONFIG.

DATE

CONC MEAS 1

SET NH4–N CONC

READING STABLE?

NH4–N:

DRIFT

K+

DRIFT

CALIBRATE

CAL. CONFIG.

DATE

CONC MEAS 1

SET K+ CONC

SET NH4–N CONC

READING STABLE?