Operation – Hach-Lange NH4D sc Ammonium Sensor User Manual

Page 27

Advertising
background image

27

Operation

8.

During the reaction phase of both cuvette/vial tests, add 40 ml
from the LCW892 mix standard to the sample in the bucket.

9.

With the probe, mix the sample in the bucket once and wait
3 minutes.

10. Select

CONC MEAS 2

and wait until the NH

4

–N and K values

are stable (drift should be <0,03 mg/l).

11. Press

ENTER

.

12. Dilute the sample for the ammonium cuvette/vial test 1:5 and

for the potassium test 1:10.

13. Perform the LCK 303 ammonium cuvette/vial test and the

LCK 228 potassium cuvette test simultaneously from the
filtered or sedimented sample.

14. When having the results of all cuvette/vial tests, enter the

ammonium and potassium concentrations for both measuring
points under

SET K+ CONC

and

SET NH4–N CONC

.

15. The message

CAL. COMPLETED OK

. appears if the calibration

was successful.

Sample concentration 3:

@ 60 mg/l K

(very high potassium concentrations in the waste water)
6.

Press

ENTER

.

7.

Perform the LCK 303 ammonium cuvette/vial test and the

LCK 228 potassium cuvette test simultaneously from the
filtered or sedimented sample (dilute the sample accordingly).

8.

During the reaction phase of both cuvette/vial tests, add 80 ml
from the LCW 892 mix standard to the sample in the bucket.

9.

With the probe, mix the sample in the bucket once and wait
3 minutes.

10. Select

CONC MEAS 2

and wait until the NH

4

–N and K values

are stable (drift should be <0,03 mg/l).

11. Press

ENTER

.

12. Dilute the sample for the ammonium cuvette/vial test 1:10 and

for the potassium test 1:25.

13. Perform the LCK 303 ammonium cuvette/vial test and the

LCK 228 potassium cuvette test simultaneously from the
filtered or sedimented sample.

14. When having the results of all cuvette/vial tests, enter the

ammonium and potassium concentrations for both measuring
points under

SET K+ CONC

and

SET NH4–N CONC

.

15. The message

CAL COMPLETED OK

. appears if the calibration

was successful.

Note: In case of a measuring range exceedance an appropriate dilution of
the sample is required.