Troubleshooting

6.3.2 Troubleshooting during calibration

Symptom

Possible cause

Corrective actions

SENSOR CODE

Sensor code entered incorrectly

Using the certificate, check whether the
sensor code was entered correctly. If not
havin

g

the certificate, carry out "Factory

Calibration" (Factory cal).

CALIBRATING

Data transfer from software 1.3x has failed

Carry out a completely new calibration: a)
sensor code or factory calibration, b) suitable
matrix correction

AMMONIUM

OFFSET

Error in the last ammonium calibration, sensor
cartridge too old, contaminated, faulty

Repeat the calibration.

Use the previous calibration.

Clean or replace the sensor cartridge.

SLOPE

Error in the last ammonium calibration, sensor
cartridge too old, contaminated, faulty

Repeat the calibration.

Use the previous calibration.

Clean or replace the sensor cartridge.

POTASSIUM

OFFSET

Error in the last potassium calibration, sensor
cartridge too old, contaminated, faulty

Repeat the calibration.

Use the previous calibration.

Clean or replace the sensor cartridge.

SLOPE

Error in the last potassium calibration, sensor
cartridge too old, contaminated, faulty

Repeat the calibration.

Use the previous calibration.

Clean or replace the sensor cartridge.