4 matx2 k+ correction, Tx2 k, Operation – Hach-Lange NH4D sc Ammonium Sensor User Manual

Page 26

Advertising
background image

26

Operation

10. By confirming the second laboratory value, the Matrix2

correction is activated.

4.6.3.4 MATX2 K+ correction

Select

MATX2 K+

to calibrate the ammonium and potassium sensor

simultaneously at two different measuring points.

The calibration is carried out via a standard addition in the sample
matrix. Proceed as follows to perform the MATX2 K+:

1.

Fill a 10-liter bucket with 3 to 5 liters of sample from the sample
flow in which the sensor is immersed.

2.

Then place the probe in the bucket containing the sample.

3.

Select

SENSOR SETUP>NH4D SC>CALIBRATE> CAL CONFIG

.

4.

Select

MATX2 K+

in the selection window and press

ENTER

.

5.

Select

CONC MEAS 1

and wait until the NH

4

–N and K values

are stable (drift should be <0,03 mg/l).

Additional steps for various sample concentrations:

Sample concentration 1:

" 10 mg/l NH

4

–N

and

60 mg/l K

6.

Press

ENTER

.

7.

Perform the ammonium cuvette/vial test (LCK 305 and
LCK 303) and the LCK 228 potassium cuvette test
simultaneously from the filtered or sedimented sample.

8.

During the reaction phase of both cuvette/vial tests, pipette
8 ml from the LCW892 mix standard to the sample in the
bucket.

9.

With the probe, mix the sample in the bucket and wait
3 minutes.

10. Select

CONC MEAS 2

and wait until the NH

4

–N and K values

are stable (drift should be <0,03 mg/l).

11. Press

ENTER

.

12. Dilute the sample for the cuvette/vial tests 1:3.

13. Perform the LCK 303 ammonium cuvette/vial test and the

LCK 228 potassium cuvette test simultaneously from the
filtered or sedimented sample.

14. When having the results of all cuvette/vial tests, enter the

ammonium and potassium concentrations for both measuring
points under

SET K+

and

SET NH4–N

.

15. The message

CAL. COMPLETED OK

. appears if the calibration

was successful.

Sample concentration 2: 10 100 mg/l NH

4

–N and 60 mg/l K

6.

Press

ENTER

.

7.

Perform the LCK 303 ammonium cuvette/vial test and the

LCK 228 potassium cuvette test simultaneously from the
filtered or sedimented sample (if necessary, dilute the sample
accordingly).

CALIBRATE

CAL. CONFIG.

CONC MEAS 1

DATE

SET K+ CONC

SET NH4–N CONC

CONC MEAS 2

DATE

SET K+ CONC

SET NH4–N CONC