5 value correction, Ue correction (see, Operation – Hach-Lange NH4D sc Ammonium Sensor User Manual

Page 28

Advertising
background image

28

Operation

4.6.3.5 Value correction

Value correction offers the option of subsequently correcting a
matrix at two different concentrations.

Take several samples on various days with different concentrations
and perform an analysis of the samples in the laboratory.

Note: The concentrations should be in a concentration range of at least
half a decade.

1.

Note the two displayed ammonium and potassium values,
which are measured with the sensor at the time of the
sampling.

2.

Then note the measured ammonium laboratory reference
value.

All 3 values form a correction point.

3.

From the values recorded, select the two correction points with
ammonium concentrations as far apart as possible.

4.

Select

CAL CONFIG.> CALIBRATE>VALUE CORR

from the menu

and press

ENTER

to confirm.

Enter the 3 noted values:

5.

1. NH4–N NH4D sc: Enter the displayed NH

4

–N value for the

1st correction point. Press

ENTER

to confirm.

6.

1. K+ NH4D sc: Enter the displayed K+ value for the 1st
correction point. Press

ENTER

to confirm.

7.

1. NH4–N lab: Enter the measured NH

4

–N laboratory

reference value for the 1st correction point. Press

ENTER

to

confirm.

8.

2. NH4–N NH4D sc: Enter the displayed NH

4

–N value for the

2nd correction point. Press

ENTER

to confirm.

9.

2. K+ NH4D sc: Enter the displayed K+ value for the 2nd
correction point. Press

ENTER

to confirm.

10. 2. NH4–N lab: Enter the measured NH

4

–N laboratory

reference value for the 2nd correction point. Press

ENTER

to

confirm.

Value correction is activated and displayed as

MATX2

.

Advertising