English, Specifications – Hach-Lange DR 3900 Basic User Manual User Manual

Page 103

Advertising
background image

English 103

English

Specifications

These are subject to change without notice!

Performance specifications

Operating mode

Transmittance (%), Absorbance and Concentration

Source lamp

Halogen lamp

Wavelength range

320–1100 nm

Wavelength Accuracy

± 1.5 nm (wavelength range 340–900 nm)

Wavelength
reproducibility

± 0.1 nm

Wavelength resolution

1 nm

Wavelength calibration Automatic

Wavelength selection

Automatic, based on method selection

Scanning speed

≥ 8 nm/s (in steps of 1 nm)

Spectral bandwidth

5 nm

Photometric measuring
range

± 3.0 Abs (wavelength range 340–900 nm)

Photometric accuracy

5 m Abs at 0.0–0.5 Abs
1 % at 0.50–2.0 Abs

Photometric linearity

< 0.5 % to 2 Abs
≤1 % at > 2 Abs with neutral glass at 546 nm

Stray light

< 0.1 % T at 340 nm with NaNO

2

Data log

2000 readings (Result, Date, Time, Sample ID, User
ID)

User programs

100

Physical and environmental specifications

Width

350 mm (13.78 in)

Height

151 mm (5.94 in)

Depth

255 mm (10.04 in)

Mass

4200 g (9.26 lb)

Ambient operating
requirements

10–40 °C (50–104 °F), maximum 80 % relative
humidity (without condensate formation)

Ambient storage
requirements

–40–60 °C (–40–140 °F), maximum 80 % relative
humidity (without condensate formation)

Additional technical data

Power connector via
external power supply

Input:100–240 V/50–60 Hz
Output: 15 V/40 VA

Interfaces

Use only shielded cable with maximum length of 3 m:
2× USB type A
1× USB type B
Use only shielded cable (e.g. STP, FTP, S/FTP) with
maximum length of 20 m:
1× Ethernet

Enclosure rating

IP40 (excluding interfaces and power supply)

Protection class

Class I

Performance specifications