English 119

Error during default
gateway setup.

Network setup: default
gateway cannot be set
for fixed IP address

Try to create the
connection again.

Error during network
drive setup!

Error during network
setup

Check the settings.

Error during subnet
mask setup.

Network setup: Subnet
mask cannot be set for
fixed IP address

Enter the subnet mask
again.

Error copying from USB
memory stick.

Error during update

Start the procedure again
or contact the
manufacturer or his
representative.

Error in FTP
connection.

FTP error

Make sure that the
instrument is connected
to the network.

Fault
Program not
accessible.
Please update program
data!

Barcode test not present

Data updating

Fault
Clean cuvette!

The cuvette is soiled or
there are undissolved
particles in the cuvette

Clean the cuvette; allow
the particles to settle

Fault
Test program stopped!
Please check lamp
Close the lid.
Error [xx]

Test program stops when
the instrument is started

Check the lamp and
replace it if necessary.
Close lid.
Press Start Again.

Fault
Test program stopped!
Please remove the
cuvette
Close the lid.

Test program stops when
the instrument is started

Remove the cuvette/
sample cell from the cell
compartment.
Press OK.

Error displayed

Definition

Resolution

Error
Selfcheck stopped.
Hardware error.
Error [x]

Electronic defect

Contact the
manufacturer or a sales
representative and
indicate the error number

Error
Too much ambient
light!
Move instrument into
shade
or close the lid

The instrument sensors
detects too much ambi-
ent light.

Decrease ambient light.
(Avoid direct sunlight.)
Close lid.

No help function is
available for this
program.

Shelf life exceeded!
Use chemicals?

The analysis is possibly
erroneous. Use new
chemicals

No evaluation!

Error in the test data-
base / user database

Check programming
Contact the
manufacturer or a sales
representative

No barcode!

No barcode found

Reinsert the cell.
If the barcode is not
recognised, contact the
manufacturer or his
representative.

No instrument backup
present!

Check the USB memory
stick.

No valid data for these
parameters!

Data analysis not possi-
ble, no measurement
data

Change the selection.

No valid data found!

View Data not possible in
data log

Change the selection.

No help function
present.

Error displayed

Definition

Resolution