Installation

3.3.1

PHOSPHAX sc analyzer power supply

Refer to the PHOSPHAX sc manual.

3.4

sc1000 controller connection

Connect the SUB-D plug supplied to a dual-core, sheathed data cable (signal or bus
cable). For further information regarding the data cable connection, refer to the enclosed
assembly instructions.

3.5

Flow rate signal connection

If a flow rate measurement signal of 4 to 20 mA is available, connect it to the analog input
of the RTC module.

3.6

Connection to the automation unit on the plant side

Depending on the variant and option, the RTC101 P-Module is equipped with various
components that must be connected to the plant's automation unit.

•

The volumetric flow rate is provided to the RTC module as a 0/4 to 20 mA signal for all
variants and options.

•

The precipitant volume to be metered is delivered by the RTC module as a 0/4 to
20 mA signal for all variants and options.

Alternatively, the precipitant volume can also be output by the fieldbus variants
provided by the sc1000 (see sc1000 operating instructions).

•

The RTC module supplies the digital output signal for activating the precipitant pump
at 0 V or 24 V.

•

The RTC module supplies a collective fault message at 0 V (fault) or 24 V (device
functional).

Table 2 Signal allocation of individual components of the RTC module

RTC module options

Component

Name

Connection

Signal

Function

1-channel

2-channel

2-fold digital output

1

1

Ground to connections 3 and 7 or equal to voltage supply

KL2032

1

+24 V/0 V

Precipitant pump on/off

X

5

+24 V/0 V

No fault/fault

X

4-fold digital output

1

KL2134

1

+24 V/0 V

Precipitant pump 1 on/off

X

5

+24 V/0 V

No fault/fault in channel 1

X

4

+24 V/0 V

Precipitant pump 2 on/off

X

8

+24 V/0 V

No fault/fault in channel 2

X

Single analog output

KL4011

1(+) 3(-)

+24 V/0 V

Precipitant pump dosing rate

X

2-fold analog output

KL4012

1(+) 3(-)

+24 V/0 V

Precipitant pump 1 dosing rate

X

5(+) 7(-)

+24 V/0 V

Precipitant pump 2 dosing rate

X

1-fold analog input

KL3011

1(+) – 2(-)

0/4 to 20 mA

Feed volume flow Channel 1

X

X

2-fold analog input

KL3011

1(+) – 2(-)

0/4 to 20 mA

Feed volume flow Channel 2

X