Parameterization and operation – Hach-Lange WTOS P-RTC User Manual
Page 28

28
Parameterization and operation
CONFIGURE (CONTINUE)
PRECIPITANT
CHANNEL 1
METAL CONTENT
Metal concentration in precipitant (refer to
) [g/L]
ATOMIC WEIGHT
Relative atomic weight of active precipitant substance (refer to
CHANNEL 2
METAL CONTENT
Metal concentration in precipitant (refer to
) [g/L]
ATOMIC WEIGHT
Relative atomic weight of active precipitant substance (refer to
MODBUS
ADDRESS
Start address of an RTC module within the MODBUS network. Default is 41. This
setting must only be changed by the manufacturer's service department (
).
(Refer to
)
DATA ORDER
Specifies the register order within a double word. Default is NORMAL. This setting
must only be changed by the manufacturer's service department (Refer to
DATALOG INTRVL
Indicates the interval in which the data is saved in the log file.
[min]
MAINTENANCE
RTC DATA
RTC MEASUREMEN
Displays up to 5 measured values; additional values can be selected by scrolling.
RTC ACTUAT VAR
Displays up to 5 actuating variables; additional variables can be selected by scrolling.
DIAG/TEST
EEPROM
Hardware test
RTC COMM TO
Communication time-out
RTC CRC
Communication checksum
LOCATION
A location name can be assigned here for better identification of RTC module, e.g.
activation 2.
SOFTWARE VERSION
Version number for service
RTC MODE
Indicates the mode set in the RTC module.
2-channel closed-loop control