Parameterization and operation

Figure 3

Program change

If both measurement signals become invalid at the same time,
the system switches between programs 4 and 1 without
intermediate stages.

4.2.2

Manual pre-selection

Manual pre-selection limits the selection of programs.

4.2.3

CF card configuration

The function of the RTC101 P-Module, i.e. control/regulation, is
indicated on the CF card. If this setting is to be changed, please
contact the manufacturer's service department (

Section 8

).

Q/PO

4

-P

Program 4

Q profile/PO

4

-P

Program 3

Q profile/PO

4

-P profile

Program 1

Q/PO

4

-P profile

Program 2

Pre-selection 1:

Always Program

1

Pre-selection 2:

Open-loop
control
:

Program 2

On signal drop out

Program 1

Pre-selection 3:

Closed-loop
control
:

Program 3

On signal drop out

Program 2 if possible

Program 3

Otherwise program 1

N O T I C E

Never remove the CF card from the RTC module during operation!
This can damage the instrument!