Section 6 troubleshooting, 1 error messages, 2 warnings – Hach-Lange WTOS P-RTC User Manual

Page 39: 3 wear parts, 1 error messages 6.2 warnings 6.3 wear parts

Advertising
background image

39

Section 6

Troubleshooting

6.1

Error messages

The sc controller displays the possible sensor errors.

6.2

Warnings

The sc controller displays the possible warning messages from the sensor.

6.3

Wear parts

Displayed errors

Cause

Resolution

RTC MISSING

No communication between RTC and RTC

communication card

Supply RTC with voltage
Test connection cable
Reset the sc1000 and the RTC (switch so it is

completely voltage free and switch back on)

RTC CRC

Interrupted communication between RTC and

RTC communication card

Make sure +/- connections of the connector

cable between RTC and RTC communication

card in the sc1000 are installed correctly.

CHECK CONFIG

The sensor selection of the RTC was deleted

by deleting or selecting a new sc1000

participant.

From MAIN MENU > RTC MODULES /

PROGNOSYS > RTC MODULES > RTC >

CONFIGURE > SELECT SENSOR, select the

correct sensor for the RTC again and confirm.

RTC FAILURE

Brief general read/write error on the CF card,

mostly caused by a brief interruption to the

power supply.

Acknowledge error. If this message is shown

frequently, eliminate the cause of the power

disruptions. If necessary, inform the service

team of the manufacturer (

Section 8, page 43

).

FAULTY PUMP 1

Faulty measurement signal

Test sensor, check cable connections

FAULTY PUMP 2

Faulty measurement signal

Test sensor, check cable connections

INFLUENT1 NOT G.

Faulty measurement signal

Test sensor, check cable connections

INFLUENT2 NOT G.

Faulty measurement signal

Test sensor, check cable connections

Displayed warnings

Cause

Resolution

MODBUS ADDRESS

The RTC menu

PLANT CONFIG

was opened.

This deletes the MODBUS address of the RTC
module.

OPEN THE MODBUS

menu of the RTC module

and set the correct MODBUS address.

PROBE SERVICE

A configured sensor is in service status.

The sensor must exit service status.

Designation

Number

Service life

CF card, type RTC module

1 piece

2 years

Battery, type CR2032 Panasonic or Sanyo

1 piece

5 years

Advertising