21

Parameterization and operation

2-channel open-loop control

CONFIGURE

SELECT SENSOR

Select the sensors installed for the open-loop control (see

Section 4.4, page 29

).

OPEN-LOOP

PRECIP. TYPE

Precipitation, simultaneous precipitation, post-precipitation

CHANNEL 1

SETPOINT PO4-P

Desired orthophosphate value in effluent (refer to

4.5.1, page 31

) [mg/L]

CORR FACTOR

Percentage correction of precipitant dosing (see

4.5.2, page 31

) [%]

BIO-P

Phosphate biologically eliminated after influent (refer to

4.5.3, page 31

) [%]

MIN DOS RATE

Minimum flow rate of dosing pump

[L/h]

PRESELECT PROG

Programs 1 to 4 (refer to

4.1, page 15

)

CHANNEL 2

SETPOINT PO4-P

Desired orthophosphate value in effluent (refer to

4.5.1, page 31

) [mg/L]

CORR FACTOR

Percentage correction of precipitant dosing (see

4.5.2, page 31

) [%]

BIO-P

Phosphate biologically eliminated after influent (refer to

4.5.3, page 31

) [%]

MIN DOS RATE

Minimum flow rate of dosing pump

[L/h]

PRESELECT PROG

Programs 1 to 4 (refer to

4.1, page 15

)

PROFILE

Active

when a measurement signal fails

Q-PROFILE

Daily profiles (2 h average) of waste water influent according to feed measurement
signal

[m³/h]

DIST PROFILE

Percentage distribution of flow rate; input refers to channel 1.

[%]

P-PROFILE

Daily profiles of PO

4

-P concentrations, 2 h average (see

4.5.4, page 32

) [mg/L]

WEEK PROFILE

Percentage daily averages of the phosphate load (refer to

4.5.4, page 32

) [%]

IN- OUTPUTS

DOSING PUMP

CHANNEL 1

MIN PUMP RANGE

Lower threshold of flow rate range

[L/h]

MAX PUMP RANGE

Upper threshold of flow rate range

[L/h]

0/4...20MA

Selection of transfer range according to pump input

CONTROL CYCLE

Control cycle comprising on- and off-time (see

4.5.6, page 33

)

[s]

MIN RUNTIME

Minimum on-time of pump (refer to

4.5.6, page 33

) [s]