Installation

Page 35

Manual BÜHLER 1027

Characteristic features: BÜHLER 1027 sampler

The device consists of the following component parts:

1.

Housing/protective top

2.

Suction hose connector

3.

Metering vessel

4.

Pinch Valve

5.

Discharge tube

6.

Control unit with back-lit
LC-display

7.

Keypad

8.

RS 232 interface

9.

Cable feedthrough for mains
cable

10. Cable feedthrough for air

inlet/outlet

11. Signal connector

Installing the apparatus

-

We recommend to install the apparatus as close to the sampling point as possible.

-

The hose has to be laid with a constant fall from the sampler to the sampling point as lower
lying points can lead to deposits in the hose which may freeze in winter or result in cross-
contaminations.

-

To fix the suction hose, we recommend to use the extraction unit available as accessory.

-

Immerse the end of the hose at the sampling point with the open end facing downstream
(direction of flow) so that coarse matter and fibres cannot be forced into the suction aperture

Mains connection

The device is equipped with a power supply lead.

The connecting data can be found in chapter “Specifications”.

Switch on / off

The device is switched on and off by means of the shock-proof plug.

Hose connection

The opening for connection of the suction hose is top right beside the dosing
unit.

Signal connections

The connector for signals (e.g. flow meter) is located at the bottom of the
control housing

Messages

Collective malfunction message (is directly connected to the relay).