Operation – Hach-Lange BUHLER 1027 wall mount automatic water sampler (BL series) User Manual

Page 47

Advertising
background image

Operation

Page 47

Manual Bühler 1027

5.2.5

**********************

BOTTLE VERSION
*SINGLE BOTTLE
2 x 5;0 L
2 x 10,0 L

**********************

If your sampler is equipped with a distributor, select here the
distributor/bottle variant.

5.2.12

**********************

REPEAT PROGRAM?

YES
*NO

**********************

If the program shall only run once, select NO.

Selecting „YES“ means continuous operation (the program is
always restarted).

Remark:

In that case, the program will run endlessly. So,
please make sure that the bottles are emptied
regularly according to the programming as
otherwise they will be overfilled.

1.1.4

**********************

PROGRAM NO. 2
FLOW ANAL. 0-20 mA
SAMP./H AT 20 mA 40
FILLING TIME

2:00

**********************

Each programming is completed with this menu to have the
possibility to check the program settings of the just created
program.
Program No. 2 has now been programmed with a flow-
dependent program.

To start the program, press the „DOWN“ arrow key.

1.1.5

**********************

PROGRAM NO. 2
CHANGE
DELAYED START
*START

**********************

The just created program can either be
started immediately
started with a delay or
• changed.

3.

Programming of an „EVENT-PROPORTIONAL“ sampling program

If this sampling mode is selected, the sampler is waiting for an external „event“ signal, e.g. from a
connected pH-meter. A sample is extracted according to the programming as long as the signal is
present. When the signal fades, the sampler waits for the next signal and then fills the next empty
bottle. Which „event“ sample has been filled into which bottle is recorded in the info memory.

1.1.4

**********************

PROGRAM NO. 3
TIME
INTERVAL HRS 0:06
FILLING TIME H 2:00

**********************

Selection of the program number:

After having pressed the ENT key, a program number (1 to 6) can
be entered.
Confirm by pressing the ENT key.

Next menu: press the „DOWN“ arrow key.

Advertising