Operation

Page 44

Manual Bühler 1027

3.4.1

**********************

MODE
EVENT
*TIME
FLOW ANAL. 0-20 mA

**********************

The selected sampling mode is marked with a star.
If the sampling mode shall be changed, press the ENT key and move
the cursor on the sampling mode TIME to create a time-proportional
program.

Confirm the input by pressing the ENT key.

4.12.1

**********************

INTERVAL

00:10

**********************

Enter the sampling interval (hh:mm).
A sample is e.g. extracted every 10 minutes.

5.2.2

**********************

BOTTLE FILLING

*TIME-RELATED
SAMPLE-RELATED

**********************

Select whether the bottle filling shall be

1.

time-dependent or

2.

depend on a pre-programmed number of

samples

6.2.1

**********************

FILLING TIME PER
BOTTLE

2:00H

**********************

TIME-RELATED bottle filling:
Enter for how long each bottle shall be filled until the distributor
changes to the next bottle.
Example:

Interval:

10 min.

Filling time/bottle: 120 min.
Bottle volume:

2,5 L

That means 12 samples are extracted in 120 min. With a bottle
volume of 2,5 L, the maximum dosing volume per sample is
calculated as follows:
2500:12

200 ml.

5.2.5

**********************

BOTTLE VERSION
*SINGLE BOTTLE
2 x 5;0 L
2 x 10,0 L

**********************

If your sampler is equipped with a distributor, select here the
distributor/bottle variant.

5.2.12

**********************

REPEAT PROGRAM

YES
*NO

**********************

If the program shall only run once, select NO.

Selecting „YES“ means continuous operation (the program is
always restarted).

Remark:

In that case, the program will run endlessly. So, please
make sure that the bottles are emptied regularly
according to the programming as otherwise they will be
overfilled.