Page 48

Manual Bühler 1027

1.1.5

**********************

PROGRAM NO. 3
*CHANGE
DELAYED START
START

**********************

PROGRAMMING

- Press the ENT key (cursor is displayed flashing)
- Press the „UP“ arrow key twice
- Confirm with ENT

2.2.1

**********************

PASSWORD

****

**********************

Enter the valid password (factory-set code is 6299).

Remark:
This is only required if a password has been activated in „Setup“->
„System settings“. If not, just press the ENT key and you will get to
the next menu!

If the right password has been entered, you will automatically get
to the next menu.

3.4.1

**********************

MODE
*EVENT
TIME
FLOW 0-20 mA

**********************

The selected sampling mode is marked with a star.

If the sampling mode shall be changed, press the ENT key and
move the cursor on the sampling mode EVENT to create an
event-related sampling program.

Confirm the input by pressing the ENT key.

4.12.1

**********************

INTERVAL

00:10

**********************

Enter the sampling interval (hh:mm).
A sample is e.g. extracted every 10 minutes.

5.2.2

**********************

BOTTLE FILLING

*TIME-RELATED
SAMPLE-RELATED

**********************

Select whether the bottle filling shall be

1. time-dependent or
2. depend on a preprogrammed number of samples.

6.2.1

**********************

FILLING TIME PER
BOTTLE

2:00H

**********************

At TIME-RELATED bottle filling:
Enter for how long each bottle shall be filled until the distributor
changes to the next bottle.
Example:

Interval:

10 min.

Filling time/bottle: 120 min.
Bottle volume:

2,5 L

That means 12 samples are extracted in 120 min. With a
bottle volume of 2,5 L, the maximum dosing volume per
2500:12

200 ml.