Operation
Page 46
Manual Bühler 1027
3.4.1
**********************
MODE
EVENT
TIME
*FLOW 0-20 mA
**********************
The selected sampling mode is marked with a star.
If the sampling mode shall be changed, press the ENT key and
move the cursor on the sampling mode FLOW to create a flow-
dependent sampling program.
Confirm the input by pressing the ENT key.
4.10.1
**********************
SAMPLES PER HOUR
AT 20 mA
40
**********************
If „ANALOGUE“ (0/4-20mA) has been selected, enter here how
often samples shall be extracted at a maximum analogue flow
signal.
Example:
- Flow analogue 0-20 mA
- At 20 mA 40 samples/hour shall be extracted.
- If there are on an average 10 mA for a period of
1 hour, 20 samples will be extracted.
5.2.2
**********************
BOTTLE FILLING
*TIME-RELATED
SAMPLE-RELATED
**********************
Select whether the bottle filling shall be
1. time-dependent or
2. depend on a pre-programmed number of samples.
6.2.1
**********************
FILLING TIME PER
BOTTLE
2:00H
**********************
At TIME-RELATED bottle filling:
Enter for how long each bottle shall be filled until the distributor
changes to the next bottle.
Example:
Interval:
10 min.
Filling time/bottle:
120 min.
Bottle volume:
2,5 L
That means 12 samples are extracted in 120 min. With a bottle
volume of 2,5 L, the maximum dosing volume per sample is
calculated as follows: 2500:12
200 ml.
5.2.3
**********************
NO. OF SAMPLES PER
BOTTLE LIMITED?
*YES
NO
**********************
In addition to the entry of a bottle filling time you can also limit the
number of samples per bottle.
6.5.1
**********************
MAX. NO. OF SAMPLES
PER BOTTLE
5
**********************
Enter the maximum number of samples per bottle.