Operation – Hach-Lange BUHLER 1027 wall mount automatic water sampler (BL series) User Manual

Page 46

Advertising
background image

Operation

Page 46

Manual Bühler 1027

3.4.1

**********************

MODE
EVENT
TIME
*FLOW 0-20 mA

**********************

The selected sampling mode is marked with a star.
If the sampling mode shall be changed, press the ENT key and
move the cursor on the sampling mode FLOW to create a flow-
dependent sampling program.

Confirm the input by pressing the ENT key.

4.10.1

**********************

SAMPLES PER HOUR
AT 20 mA

40

**********************

If „ANALOGUE“ (0/4-20mA) has been selected, enter here how
often samples shall be extracted at a maximum analogue flow
signal.
Example:

- Flow analogue 0-20 mA
- At 20 mA 40 samples/hour shall be extracted.
- If there are on an average 10 mA for a period of

1 hour, 20 samples will be extracted.

5.2.2

**********************

BOTTLE FILLING

*TIME-RELATED
SAMPLE-RELATED

**********************

Select whether the bottle filling shall be

1. time-dependent or
2. depend on a pre-programmed number of samples.

6.2.1

**********************

FILLING TIME PER
BOTTLE

2:00H

**********************

At TIME-RELATED bottle filling:
Enter for how long each bottle shall be filled until the distributor
changes to the next bottle.
Example:
Interval:

10 min.

Filling time/bottle:

120 min.

Bottle volume:

2,5 L

That means 12 samples are extracted in 120 min. With a bottle
volume of 2,5 L, the maximum dosing volume per sample is
calculated as follows: 2500:12

200 ml.

5.2.3

**********************

NO. OF SAMPLES PER
BOTTLE LIMITED?
*YES
NO

**********************

In addition to the entry of a bottle filling time you can also limit the
number of samples per bottle.

6.5.1

**********************

MAX. NO. OF SAMPLES
PER BOTTLE

5

**********************

Enter the maximum number of samples per bottle.