19

Standard Operations

5.5

Cell menu

Press three times the RIGHT arrow key in the Main
menu to move to the Electrode menu.

Figure 8 Cell menu

Apply internal stirring

1.

Select Internal stirring = ON,

2.

In the field Speed, select a stirring speed.

Apply external stirring

1.

Connect the Stirring Propeller, part no. 904-480,

to the Titration Manager (see

Figure 2 on page

10

, position 14).

Line 1 is automatically replaced by External
stirring
.

2.

Select External stirring = ON.

Question mark:
There is a problem in the editing of the electrode system. You need to be in Supervisor
mode to solve the problem. Check the sequence/method parameters of the reagent or
the electrode.
Check the sequence or method, (press 1 in the Main window). The instrument indicates
the possible errors and prompts you to correct them until ? disappears.

Table 2 Electrode icons

Icon

Description

1

Title bar: indicates the name of the window.

3

Speed: select the internal stirring speed, from
100 to 1100 rpm by steps of 50 rpm.

2

Internal stirring: command stirrer On/Off (see

Apply internal stirring

).

4

Cell icon: animated icon indicates when the
magnetic stirrer or propeller is operating.

1

3

2

4