8

Section 3

Installation

3.1

Unpacking the instruments

The BIOGAS Titration Manager comes packaged with the following items:

BIOGAS Titration Manager, one burette, zero version

B525, Burette Stand, 25 mL, red cover with connection tubing

Accessories for Titration Manager burette stand

Set of Cables for Tiration Workstation (w/o serial PC Printer cable)

Cell kit

Titration Vessel, PP, 40-100 mL (pack of 10 pcs of 956-265)

Titration Vessel, PP, 22-45 mL (pack of 10 pcs of 956-178)

User Manual BIOGAS

Basic User Manual BIOGAS, multilingual

Method Guide for FOS/TAC (organic acids (FOS) and total inorganic carbon (TAC)), En-Ger

Installation form (Fr-Eng-Ger)

Note: If any of these items are missing or damaged, contact the manufacturer or a sales
representative immediatly.

3.2

Operating environment

The following conditions are necessary to ensure correct instrument operation and accurate results:

Place the instrument firmly on an even surface. Do not push any objects under the instrument.

Maintain an ambient temperature of 5 to 40 ºC (41 to 104 ºF) for proper instrument operation.

The relative humidity should be less than 80 %; moisture should not condense on the
instrument.

Leave at least a 15 cm (6 in.) clearance at all sides for air circulation to avoid overheating of
electrical parts.

Do not operate or store the instrument in extremely dusty, damp or wet locations.

Keep the surface of the instrument and all accessories clean and dry at all times. Splashes or
spills on the instrument should be cleaned up immediately.

D A N G E R

Only qualified personnel should conduct the tasks described in this section of the manual.

N O T I C E

Protect the instrument from temperature extremes, including heaters, direct sunlight and other heat sources.