Standard Operations

5.10.3.1 Edit a method

1.

Enter the Method menu and press 5 (Method

library).

2.

Press the validation key and select the method

to be edited from the list.

3.

Press 2 Edit method.

Note: The RIGHT and LEFT arrow keys allow you to
move from one screen to the other.

4.

Select the method mode. The modes available

are:

Measurement (pH, mV or mV at imposed

current)

End point (titration to a preset mV or pH

end point)

Coupled (methods chained in a same bea-

ker)

Continuous IP (Inflection point detection

with continuous addition of the titrant)

Monotonic IP (Inflection point detection

with incremental addition of the titrant,

increments of equal size)

Dynamic IP (Inflection point detection with

incremental addition of the titrant, incre-

ments of varying size)

5.

Select the measurement mode:

pH

mV

mV at imposed current.

6.

Select Temperature.

If a temperature sensor is in use, select

Probe.

If the temperature is entered manually via

the numeric keypad, select Entered.

If the calibration is to be performed at 25°C,

select Fixed 25°C.

7.

Enter the Number of tests, i.e. the number of

times the method is to be repeated.

8.

Select the way the cell grounding of the

measuring electrode will be performed:

Reference: reference electrode connected

to the Ref socket on the titration system.

Metal: metal electrode connected to the

GND socket on the titration system. For tit-

rations in very resistant medium.

Other: an electrode which does not belong

to the electrode system.

9.

If you want a message to be displayed upon

starting the method, select Notification = Yes

then enter the message (3 lines of 32

characters maximum).