Specifications – Hach-Lange BIOGAS User Manual User Manual

Page 5

Advertising
background image

5

Specifications

Burette

1 embedded burette stand
Burette volumes available

1, 5, 10, 25, 50 mL

Burette motor

18000 steps

Complies with ISO/FDIS 8655-3
UV-protected encapsulated glass syringe

Embedded operating procedures

burette exchange, air bubble removal (Flush)
Fill, Empty function

Inputs/outputs

1 indicator electrode inputs
1 reference electrode input
Selectable polarised input from -1 mA to +1 mA in 1 µA
steps, DC or AC
Differential input
Temperature input
0-5 V TTL output

Serial connections

Printer/PC, balance, sample changer and additional
titrator for use with TitraMaster 85 PC Software

PS/2 port for PC keyboard and/or barcode reader

General specifications

Casing

Fully splashproof.
Graphic 128 x 128 dot LCD protected from spillages.
Soft-touch alphanumeric keypad (silicone)

Languages

English, German, Danish, French, Italian, Spanish,
Swedish

Dimensions (H x W x D)

380 x 230 x 450 mm (1.25 x 0.75 x 1.48 ft) (excl. tubing)

Weight

5 kg (excluding reagent bottles)

Power requirements

115 VAC (±10 %) or 230 Vac (±10 %)
50...60 Hz - 45 VA
On/Off power switch

Mains fuses

T1A L250V, slow blow (2 fuses)

Secondary fuses

T4 A, 230 V (1 fuse)
The secondary fuse is mounted on the printed circuit
board. If necessary, contact a manufacturer‘s Service
engineer for replacement of the fuse, as the instrument
casing must be opened.

Environmental operating conditions

Environmental operating conditions (working)

5 to 40 °C (41 to 104 °F), max. 80 % relative humidity
(without water condensation)

Environmental operating conditions (storage)

-20 to 60 °C (-4 to 140 °F), max. 80 % relative humidity
(without water condensation)

Altitude

 2000 m

Level of pollution: 2
Transitory overvoltage: class II

Limited Warranty

The warranty period for instruments is 2 years from the date of shipment.

Advertising