40

Chapter 2

Sunny icon

. The calibration has been performed on all the

electrodes present in the system. Everything is just right!

Cloudy icon

. The electrode calibration of one of the

electrodes present in the system should be performed within
12 or 24 hours.

Stormy icon

. The calibration date has elapsed for one of the

electrodes present in the system.

If acceptance limits have been set for the calibration: at
least one calibration result lies outside the programmed
acceptance limits.

At least one of the electrodes present in the system has not
been installed.

Check the sequence or method, (press 1 in the Main
window). The instrument prompts you to do the necessary
operations. You are guided step by step.

Question mark

. There is a problem in the editing of the

electrode system. You need to be in Supervisor mode to
solve the problem. Check the sequence/method
parameters of the reagent or the electrode.

Check the sequence or method, (press 1 in the Main
window). The instrument indicates the possible errors and
prompts you to correct them until ? disappears.

Calculator icon

. Reprocessing mode (Working mode =

Reprocessing) is set on the instrument. This icon is
displayed irrespective of the status of the electrode system
used.