Chapter 2

Apply internal stirring

Apply internal stirring

Apply internal stirring

Apply internal stirring

Apply internal stirring

1.

For Instrument, select the stirrer of the Titration Manager or the
stirrer of an ABU52 connected to the Titration Manager.

2.

Select Internal stirring = ON,

3.

Press

✓ in the field Speed and select a stirring speed.

Apply external stirring

Apply external stirring

Apply external stirring

Apply external stirring

Apply external stirring

1.

Connect the Stirring Propeller, part no. 847-731, to the Titration
Manager or ABU52 Propeller socket. Line 3 is automatically
replaced by External stirring.

2.

For Instrument, select the stirrer of the Titration Manager or the
stirrer of an ABU52 connected to the Titration Manager.

3.

Select External stirring = ON.

4.

Adjust stirring by turning the stirring propeller knob. You can consult
the corresponding table between the position (1 to 9) and the stirring
speed by pressing

✓ in the field Speed setting.

Fill the Karl-Fischer cell (TIM880 only)

Fill the Karl-Fischer cell (TIM880 only)

Fill the Karl-Fischer cell (TIM880 only)

Fill the Karl-Fischer cell (TIM880 only)

Fill the Karl-Fischer cell (TIM880 only)

1.

For Instrument, select TIM.

2.

Press 1 (Fill cell).

3.

Press the TIM880 filling cell button

then press

✓.

Empty the Karl-Fischer cell (TIM880 only)

Empty the Karl-Fischer cell (TIM880 only)

Empty the Karl-Fischer cell (TIM880 only)

Empty the Karl-Fischer cell (TIM880 only)

Empty the Karl-Fischer cell (TIM880 only)

1.

For Instrument, select TIM.

2.

Press 2 (Empty cell).

3.

Press the TIM880 emptying cell button

then press

✓.