Run electrode calibration, Run reagent calibration – Hach-Lange TITRALAB Users Guide User Manual

Page 68

Advertising
background image

68

Chapter 4

4.

Connect/Check reagent(s).

Refer to Short-Form Reminder no. 2

.

5.

Run the analysis when Sunny icons are visible in the electrode and
reagent tabs.

6.

Run an electrode calibration when a Stormy icon appears in the
electrode tab.

7.

Run a reagent calibration when a Stormy icon appears in the
titrant tab.

8.

If you are unable to display the "Run" command due to the presence
of Cloudy/Stormy icons in the Reagent and Electrode windows,
activate the "Check" command. The Titration Manager will
automatically guide you through the necessary operations required
to solve the problem(s).

Run electrode calibration

Run electrode calibration

Run electrode calibration

Run electrode calibration

Run electrode calibration

Refer to Short-Form Reminder no. 4

.

1.

Select the method.

2.

Install the electrode group.

3.

Enter the Electrodes window.

4.

Select 1 Calibrate electrode. Press

✓ to select the

electrode to be calibrated from the proposed list.

5.

Press 1 to start the calibration.

Run reagent calibration

Run reagent calibration

Run reagent calibration

Run reagent calibration

Run reagent calibration

Refer to Short-Form Reminder no. 5

.

1.

Select the method.

2.

Install the reagent to be calibrated.

3.

Install the electrode group to be used to perform the calibration.

4.

Enter the Reagents window.

5.

Select 1 Calibrate/Enter titre. Press

✓ and select the

reagent from the proposed list.

6.

Press 1 to start the calibration.