64

Chapter 3

•

Monotonic IP (Inflection point detection with incremental addition
of the titrant, increments of equal size)

•

Dynamic IP (Inflection point detection with incremental addition
of the titrant, increments of varying size)

•

Karl-Fischer

•

Stat pH/mV (using modes Stat-PID, Proportional Integral and
Derivative algorithm, Stat AAA, Adaptive Addition Algorithm and
Stat °C)

•

Addition (1 to 3 additions of reagent. No measurements. Additions
are performed simultaneously or successively)

b.

Select the measurement mode: pH, mV or mV at imposed current,
ISE direct, ISE standard addition or conductivity

c.

Select the way the temperature is to be measured.

•

Probe: temperature sensor is required during measurements.

•

Entered: temperature is to be entered manually using the
keypad.

•

Fixed 25°C: measurements are assumed to be performed
at 25°C.

d.

Enter the Number of tests, i.e. the number of times the
method is to be repeated.

e.

If you want a message to be displayed upon starting the method,
select Notification = Yes then enter the message (3
lines of 32 characters maximum).

f.

Select the way the cell grounding of the measuring electrode will be
performed:

•

Reference: reference electrode connected to the Ref
socket on the titration system.

•

Metal: metal electrode connected to the GND socket on the
titration system. For titrations in very resistant medium.

•

Other: an electrode which does not belong to the electrode
system.