Chapter 3

Monotonic IP (Inflection point detection with incremental addition
of the titrant, increments of equal size)

Dynamic IP (Inflection point detection with incremental addition
of the titrant, increments of varying size)

Karl-Fischer

Stat pH/mV (using modes Stat-PID, Proportional Integral and
Derivative algorithm, Stat AAA, Adaptive Addition Algorithm and
Stat °C)

Addition (1 to 3 additions of reagent. No measurements. Additions
are performed simultaneously or successively)

b.

Select the measurement mode: pH, mV or mV at imposed current,
ISE direct, ISE standard addition or conductivity

c.

Select the way the temperature is to be measured.

Probe: temperature sensor is required during measurements.

Entered: temperature is to be entered manually using the
keypad.

Fixed 25°C: measurements are assumed to be performed
at 25°C.

d.

Enter the Number of tests, i.e. the number of times the
method is to be repeated.

e.

If you want a message to be displayed upon starting the method,
select Notification = Yes then enter the message (3
lines of 32 characters maximum).

f.

Select the way the cell grounding of the measuring electrode will be
performed:

Reference: reference electrode connected to the Ref
socket on the titration system.

Metal: metal electrode connected to the GND socket on the
titration system. For titrations in very resistant medium.

Other: an electrode which does not belong to the electrode
system.