Nova-Tech CBK Series User Manual

Page 16

Advertising
background image

© Adam Equipment Company 2007

14

3. If Limit Password is enabled then enter the password which will allow

you to change the limits or the operation of the beeper or bargraph.

10.5.1

Setting up check-weighing while weighing

• Press the [Low Limit] key. It will show the current low limit. The

“LO” symbol will appear on the display.

• Press the [CE] key to clear the old value and then enter the new

low limit using the numeric keys. The decimal point is fixed at the
position that is used for the current weighing unit. When the
desired value is shown press

[Tare] to accept the value. If you

want to reset the value to zero, press

[CE] to clear the value.

• The limits are displayed in the weighing unit in use. If the

weighing unit is pounds:ounces, the limits are entered in pounds
and decimal parts of pounds. i.e. 6,0125 lb.

• To set the high limit press [High Limit], the display will show the

high limit, the “

HI” symbol will be on to the left side of the display.

Set the high limit in the same way the low limit was set.

• Pressing the [Tare] key will return the scale to weighing, with the

Check-weighing function enabled.

10.5.2

Setting up check-weighing while parts counting or % weighing

During parts counting and percent weighing the limits are set in the same
way as above. The limits are displayed in

pcs or %.

See Section 10.4 for the description of parts counting and Section 10.7 for
percent weighing.

NOTE:

1. The weight must be greater than 20 scale divisions for the check-

weighing to operate.


2. To disable the check weighing function, enter zero into both the

limits as described above. When the current limits are shown,
press

[CE] to clear the settings, enter zero and then press [Tare]

to store the zero values.

Advertising