Gentec-EO QE and XLE4 User Manual

Page 11

Advertising
background image

QE Series Joulemeter Instruction Manual Version 3.0

April 2014

11

Footnotes

Model

QE50SP-S-MB
QE50SP-H-MB

QE50LP-S-MB
QE50LP-H-MB

Optical Absorber

MB

Spectral Range

1, 8

0.19

– 20 µm

Calibrated Spectral Range
(Optional)
(QE-QED Series)

1, 8

0.248

– 2.5 µm

10.6 µm

0.3

– 2.1 µm

Typical Sensitivity

2, 7

3 V/J

Calibration Uncertainty

7, 2, 8, 9, 11

± 3%

Repeatability

< 0.5 %

Max. Pulse Energy 1.064 µm

0.266 µm

With QED @ 1.064 µm

QED @ 0.266 µm

3, 12

15 J

12.5 J

85 J
22 J

Noise Equivalent Energy (NEE)
(Typ)

2, 11

10 µJ

Max. Repetition
Rate

2, 5, 11

500 Hz

200 Hz

Typical Rise Time
(0-100%)

2, 11

300 µsec

900 µsec

Max. Pulse Width (Typ)

2, 6, 10, 11

225 µsec

675 µsec

Max. Energy Density

12

600 mJ/cm

2

@ 1.064µm, 7ns, 10 Hz

500 mJ/cm

2

@ 266nm, 7ns, 10 Hz

Max. Energy Density with QED

16 J/cm

2

@1064nm, 7nsec, Single shot

8 J/cm

2

@1064nm, 7nsec, 10Hz

6 J/cm

2

@532nm, 7nsec, 10Hz

1 J/cm

2

@266nm, 7nsec, 10Hz

Max. Average Power

Detector Alone (QE50xP-S-MB):

With optional Heatsink (QE50xP-H-MB):

12, 13

10 W (25 W with QED)
20 W (45 W with QED)

Max. Power Density

Detector Alone (QE50xP-S-MB):

With optional Heatsink (QE50xP-H-MB):
With QED :

10 W/cm² @ 10 W

5 W/cm² @ 20 W

600 W/cm²

Dimensions (H x W x D)

Detector Alone (QE50xP-S-MB):

With optional Heatsink (QE50xP-H-MB):

75 x 75 x 15 mm
75 x 75 x 44 mm

Weight

Detector Alone (QE50xP-S-MB):

With Optional Heatsink (QE50xP-H-MB):

209 g
338 g

Aperture Size

QE 50:

QED 50:

50 x 50 mm
47 x 47 mm

Aperture Area Size

QE 50:

QED 50:

25 cm

2

22.09 cm

2

Advertising