DAMAGE TO THE OPTICAL ABSORBER MATERIALS

In any time, the beam’s incident area should not be less than 10% of the detector’s aperture.
Please contact Gentec-EO to make measurements with such smaller beams.

Damage is usually caused by exceeding the manufacturer's

’pecified maximum incident:


- Average Power Density
- Peak Pulse Power Density
- Single Pulse Energy Density

Refer to the QE Series joulemeter specifications pages. This damage can also be caused when
using a detector with a contaminated absorber or attenuator surface.

The quoted damage thresholds in the specifications section refer to a visible alteration of the
absorber surface

10

. In practice a slight alteration will not affect the joulemeter response. Consider

the joulemeter to be damaged and/or out of calibration when large-scale damage is evident or you
can see the metal electrode beneath the coating

11

.


For a QED Attenuator/Diffuser

12

mounted on a QE series joulemeter, consider the detector to be

damaged and/or out of calibration

11

:

- In the presence of an optically eroded front optical component or in the presence of sparking at

the front component, accompanied by a sharp snapping noise: this phenomenon is related to
high single pulse energy density and high peak pulse power density.

- In the presence of shattered or molten optical components: this phenomenon is related to high

average power density.


- In the presence of a damaged absorber (see above).

In the case of a TEMoo (Gaussian) beam, the maximum peak power and energy density can be
calculated using the following equation:


Density (power or energy)

2Io

W

2

Where Io is the total beam power or energy

W is the beam radius at 1/e

2

and

= 3.1416

10

For QE Series detectors, the use of the appropriate “QE series Test Target " is suggested in order to
insure that the laser beam will not damage the detector’s absorber coating; contact Gentec-EO for
further instructions.

11

Contact Gentec-EO for evaluation, repair, recalibration, or replacement (refer to the WARRANTY

instructions).

12