Specifications for xle4 – Gentec-EO QE and XLE4 User Manual

Page 23

Advertising
background image

QE Series Joulemeter Instruction Manual Version 3.0

April 2014

23

1.3.5 Specifications for XLE4

Footnotes

Model

XLE4

Optical Absorber

XT : Metallic

Spectral Range

6

0.19

– 20 µm

Calibrated Spectral Range

6

0.35

– 2.5 µm

Typical Sensitivity

1, 5

1100 V/J

Calibration Uncertainty

1, 5, 6, 7, 9

± 4% at 1064nm

± 9% for other wavelengths

Repeatability

< 0.5 %

Max. Pulse Energy 1.064 µm

2

4 mJ

Noise Equivalent Energy (NEE)
(Typ)

1, 9

150 nJ

Max. Repetition
Rate

1, 3, 9

2000 Hz

Typical Rise Time
(0-100%)

1, 9

10 µsec

Max. Pulse Width (Typ)

1, 4, 8, 9

5 µsec

Max. Energy Density

90 mJ/cm

2

@ 1.064µm, 7ns, 10 Hz

Max. Average Power

0.4 W

Dimensions (H x W x D)

26.5 x 36.0 mm diam

Weight

130 g

Aperture Size

4.0 mm diameter

Aperture Area Size :

0.16 cm

2



FOOTNOTES FOR XLE4 SERIES SPECIFICATIONS:

1

Load capacitance must be

13 pF, excluding the supplied BNC to DB-

15 “Smart Interface” coaxial

cable.

190 Assuming 90 mJ/cm

2

@ 1.064µm,7ns laser beam; with a uniform energy distribution; energy

applied to full aperture. Increasing the pulse width increases the maximum measurable energy.

190 At constant power.

4

To maintain indicated calibration, sensitivity must be de-rated for longer pulses.

5

Calibrated at: 400 µJ, 10 Hz, @1.064 µm, 7 ns pulse (FWHM), semi-Gaussian beam profile, energy

applied to 80% of aperture, loaded into 1 M

/ 13 pfd.

6

For values other than calibration wavelength a typical value is recommended but not traceable to NIST.

7

Excludes non-linearities.

8

Duration at base of pulse. Divide by 2 for FWHM (Full Width at Half Maximum) duration.

9

Loaded into

1 M

/ 13 pfd.