Gentec-EO QE and XLE4 User Manual

Page 16

Advertising
background image

QE Series Joulemeter Instruction Manual Version 3.0

April 2014

16

Footnotes

Model

QE12SP-S-MT
QE12SP-H-MT

Optical Absorber

MT

Spectral Range

1, 7, 14

0.19

– 20 µm

Calibrated Spectral Range
(QE-QED Series)

1, 7

0.248

– 2.5 µm

0.3

– 2.1 µm

Typical Sensitivity

2, 13

100 V/J

Calibration Uncertainty

2, 7, 8, 10,

13

± 3%

Repeatability

< 0.5 %

Max. Pulse Energy 1.064 µm

0.266 µm

With QED @ 1.064 µm

QED @ 0.266 µm

3, 11

0.70 J
0.10 J

1.6 J

0.25 J

Noise Equivalent Energy (NEE)
(Typ)

2, 10

0.8 µJ

Max. Repetition
Rate

2, 4, 10

MAESTRO,S-LINK, M-LINK: 6000 Hz

INTEGRA: 5200Hz

Typical Rise Time
(0-100%)

2, 10

20 µsec

Max. Pulse Width (Typ)

2, 5, 9, 10

10 µsec

Max. Energy Density

11

500 mJ/cm

2

@ 1.064µm, 7ns, 10 Hz

70 mJ/cm

2

@ 532nm, 7ns, 10 Hz

70 mJ/cm

2

@ 266nm, 7ns, 10 Hz

Max. Energy Density with QED

4 J/cm

2

@1064nm, 7nsec, Single shot

2 J/cm

2

@1064nm, 7nsec, 10Hz

0.35 J/cm

2

@532nm, 7nsec, 10Hz

0.3 J/cm

2

@266nm, 7nsec, 10Hz

Max. Average Power

Detector Alone (QE12SP-S-MT):

With optional Heatsink (QE12SP-H-MT):

11, 12

3 W (7.5 W with QED)

5 W (12.5 W with QED)

Max. Power Density

Detector Alone (QE12SP-S-MT):

With optional Heatsink (QE12SP-H-MT):
With QED :

10 W/cm² @ 3 W
10 W/cm² @ 5 W

600 W/cm²

Dimensions (H x W x D)

Detector Alone (QE12SP-S-MT):

With optional Heatsink (QE12SP-H-MT):

36 x 36 x 14 mm
36 x 36 x 33 mm

Weight

Detector Alone (QE12SP-S-MT):

With Optional Heatsink (QE12SP-H-MT):

87 g

117 g

Aperture Size

QE 12:

QED 12:

12 x 12 mm

9 x 9 mm

Aperture Area Size

QE 12:

QED 12:

1.4 cm

2

0.81 cm

2

Advertising