FOOTNOTES FOR QE12/25/50/65/95 MB SERIES SPECIFICATIONS:

1

See “Personal wavelength correction” certificate

2

Load capacitance must be

30 pF, excluding the supplied BNC to DB-

15 “Smart Interface” coaxial

cable.

3

Assuming 0.6 J/cm

2

@ 1.064µm, 7ns laser beam; with a uniform energy distribution; energy applied to

full aperture. Increasing the pulse width increases the maximum measurable energy.

4

At constant power.

5

To maintain indicated calibration, sensitivity must be de-rated for longer pulses.

6

Calibrated at: 100 mJ, 10 Hz, @1.064 µm, 150 µsec pulse (FWHM), semi-Gaussian beam profile,

energy applied to 80% of aperture, loaded into 1 M

/ 30 pfd.

7

For values other than calibrated wavelength only a typical value is available.

8

Excludes non-linearities.

9

Duration at base of pulse. Divide by 2 for FWHM (Full Width at Half Maximum) duration.

10

Loaded into

1 M

/ 30 pfd.

11

Maximum measurable energy, maximum energy density and maximum average power can be

increased by using an optional QED attenuator / diffuser.

12

Warning:

Detector body can reach 60°C at maximum powers.

13

@1.064µm, CW