Optional accessories, Qed attenuator / diffuser – Gentec-EO QE and XLE4 User Manual

Page 32

Advertising
background image

QE Series Joulemeter Instruction Manual Version 3.0

April 2014

32

4

OPTIONAL ACCESSORIES


4.1 QED ATTENUATOR / DIFFUSER

The QED

attenuators increase the energy, energy density, average power and average power

density capabilities of the QE Series.

They are engineered to typically transmit 30-50% of the incident radiation to the detector in a near
Lambertian pattern (very wide diffusion pattern).

They feature ease of installation and removal.
The QED attenuators can be optionally calibrated @1.064µm when purchased at the same time as
a corresponding QE joulemeter.

Fig. 4-1, QED ATTENUATOR SPECIFICATIONS



Attenuator

QED12

QED25

QED50

QED65

QED95

Spectral Range

0.19 to 2.5 µm

Optional Calibration

Spectral Range

0.3 to 2.1 µm

Typical Reflectance

40 - 50%

Max. Energy Density

16 J/cm

2

@1064nm, 7nsec, Single shot

8 J/cm

2

@1064nm, 7nsec, 10Hz

6 J/cm2 @532nm, 7nsec, 10Hz

1 J/cm

2

@266nm, 7nsec, 10Hz

Dimensions

(Lx W x D, mm)

30.5 x 41 x

12.5

44 x 55 x 12.5

69 x 80 x 12.5 85 x 97 x 12.5 115 x 127 x 12.5

For use with

QE12

QE25

QE50

QE65

QE95






Advertising