Probes
Teledyne LeCroy provides a passive probe for each WaveAce oscilloscope
channel.
Passive probes must be compensated to flatten overshoot. This is
accomplished by means of a trimmer on the probe body.
1. Attach the connector end of your probe to any channel.
2. Connect the probe end to the CAL output connector at the front of the
oscilloscope.
3. Adjust the trim pot on the probe body until the square wave is as flat as
possible.
4. Set the Probe option attenuation to 10X by pressing the Channel button
and then the Probe menu button.
5. Set the switch to 10X on the probe.
6. Attach the probe tip to the PROBE COMP~3V connector and the
reference lead to the PROBE COMP Ground connector. Press the
corresponding channel button, and then push the AUTO button.
7. Check the shape of the displayed waveform as follows:
Overcompensated
Correctly Compensated
Undercompensated
See Setting Probe Attenuation (on page 26) in the Vertical Settings and
Channel Controls section for more information.