WaveAce 1000/2000

42

922131-00 Rev A

Cursor Placement

1. Press the CURSORS front panel button. The Cursor menu opens.

2. Cursors are placed according to the cursor measurement Mode -

Manual, Track, or Auto – chosen in one of the following manners:

Manual. Set the mode to Manual by pressing the Mode option button
until Manual is selected. Press the Type option button and select
Voltage (horizontal cursors) or Time (vertical cursors). Press the Source
option button and select a source of CH1, CH2 (also, CH3 and 4 on 4
channel models). Manual Mode also allows you to choose MATH, REFA,
or REFB as a Cursor source. Press the CurA or CurB option button and
turn the Adjust front panel knob to adjust the cursors.

Track. Set the mode to Track by pressing the Mode option button until
Track is selected. Press the Cursor A or Cursor B option button and
select a source of CH1, CH2 (also, CH3 and 4 on 4 channel models) or
NONE. Press the CurA or CurB option button and turn the Adjust front
panel knob to adjust the cursors. If Track is selected, both cursors move
in unison and appear slightly brighter on the grid display.

Auto. Set the mode to Auto by pressing the Mode option button until
Auto is selected. When you select measurement parameters using the
MEASURE front panel button, the cursors are automatically displayed.

NOTE: When using Cursors, their measurement values are always shown in the
upper-left corner of the grid display. The cursor selected for placement is
indicated by a highlighted icon on the menu and the cursor itself has a higher
brightness on the grid display.