Teledyne LeCroy QPHY-HDMI User Manual

Page 33

Advertising
background image

QPHY-HDMI Operator’s Manual

10. Save Channel X to M1.

• Click FileSave Waveform.
• Set Save To Memory.
• Set the Source to CX.
• Set the Destination to M1.
• Click Save Now.

11. Disconnect Channel X from the Fast Edge Output and connect Channel Y to the Fast

Edge Output.

12. Press the Clear Sweeps button on the front panel to reset the averaging.

13. Allow multiple acquisitions to occur until the waveform is stable on the screen.

14. From the Channel Y menu, adjust the Deskew of Channel Y until Channel Y is directly

over the M1 trace.

15. Ensure that P3 and P2 are reasonably close to the same value. (Typically < 5ps

difference)

922540 Rev A

29

Advertising