Tqsh/tqsl, dqs output high/low time, Figure 44 – tqsh/tqsl results – Teledyne LeCroy QPHY-DDR4 User Manual

Page 55

Advertising
background image

QPHY-DDR4 Software Option

tQSH/tQSL, DQS Output High/Low Time

These tests measure the high time (tQSH) and low time (tQSL) for each valid DQS transition during a R
burst. Both the maximum and minimum tQSH/tQSL values are measured.

These tests are very similar to

tQSH and tQSL, which are measured on R bursts.


Note: Only tQSH will be discussed below. The measurement methodology is the same for the tQSL
expect the measurement is made on DQS from falling edge to a rising edge instead of a rising edge
to a falling edge.

After the completion of the tQSH test the oscilloscope is in the following configuration:

Figure 43 - Oscilloscope Configuration after the tQSH test

Shown on this screen:

• Z2 is a zoom of F2, the acquired DQS signal after any probe deskew has been applied. The zoom is

positioned at the location of the “worst case” tQSH measurement indicated by t@tQSHmin. A
trace label is applied at Vref on this trace according to the signal name assigned to DQS.

Figure 44 – tQSH/tQSL Results

In the Measure section:

tQSH (P1) is measuring the high time of DQS. The high time is determined by measuring the time

DQS crosses Vref on a rising edge at to the time DQS crosses the next associated falling edge at

924291 Rev A

55