Tdqs/Tdqh, DQ to DQS Setup/Hold Time

The purpose of this test is to characterize the setup and hold time between DQ and DQS on W bursts.
The maximum value is measured for both Tdqs and Tdqh.


Note: Only Tdqs will be discussed below. The measurement methodology is the same for the Tdqh
expect the hold time is measured rather than the setup time.

After the completion of the Tdqs test the oscilloscope is in the following configuration:

Figure 61 - Oscilloscope Configuration after the Tdqs test

Shown on this screen:

• Z2 is a zoom of F2, the acquired DQS signal after any probe deskew has been applied. The zoom is

positioned at the location of the “worst case” Tdqs measurement indicated by t@Tdqsmax. A
trace label is applied at Vref on this trace according to the signal name assigned to DQS.

• Z3 is a zoom of F3, the acquired DQ signal after any probe deskew has been applied. The zoom is

positioned at the location of the “worst case” Tdqs measurement indicated by t@Tdqsmax. A
trace label is applied at Vref + (0.5 * VdIVW) on this trace according to the signal name assigned
to DQ.

Figure 62 – Tdqs/Tdqh Results

68

924291 Rev A