ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual

Page 119

Advertising
background image

Transferring Device Check and Activity Log Files Through Wi-Fi (Optional)

9650-0912-01 Rev. K

ZOLL R Series Operator’s Guide

11–9

TRANSFER FAILED
Network Error: General

Errror Number: NNNN
Contact Network Administrator

The transfer failed because of a general network
Wi-Fi error. NNNN may indicate one of the following
error codes:

N = 7004
N = 7010
N = 7011
N = 7015
N = 7016
Action: Install another Data COMM card.

N = 7500
N = 7501
N = 7502
N = 7503
N = 7504
N = 7505
N = 7506
N = 7507
Action: Install another Data COMM II card.

N = 7017
Action: Make sure that the Defib Upload Server is
running, is operational, and can be reached on the
network.

TRANSFER FAILED
Network Error: Unknown [-7002]

Action: Make sure that no illegal characters are used
for the SSID names.

TRANSFER FAILED
Wi-Fi Network Not Found
Contact Network Administrator

The R Series could not associate with a Wi-Fi
network.

Action:

1) Make sure that the configuration values for the
SSID names are correct.

2) Make sure that the R Series unit is within range of
the wireless server.

3) Contact the Network Administrator for assistance.

TRANSFER FAILED
Could Not Connect To Server
Contact Network Administrator

The R Series could not communicate with the
DefibUpload Server.

Action:
1) Make sure that the DefibUpload Server is running,
is operational, and can be reached on the network.
2) If WEP is enabled, make sure that the WEP keys
on the R Series and Access Point match (value and
slot location).
3) Make sure that the server port is correct.

TRANSFER FAILED
Network Error: DHCP
Contact Network Administrator

The DHCP request has failed.

Action:
1) If WEP is enabled, make sure that the WEP keys
on the R Series and Access Point match (value and
slot location).
2) Make sure that the DHCP server is operational.

Advertising