Ac charger cpr – ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual

Page 139

Advertising
background image

AC Charger

9650-0912-01 Rev. K

ZOLL R Series Operator’s Guide

13–7

AC Charger

CPR

11. No 30J TEST OK message is

displayed when performing a
manual defibrillator self-test.

Check to make sure unit is set to 30 joules.

If testing with OneStep electrodes, make sure that the
OneStep cable patient connector is firmly inserted into
the electrode connector. Ensure the OneStep electrode
package is sealed.

If testing with the Test Port, make sure the OneStep
cable is firmly inserted into the Test Port.

If testing with paddles, make sure to press the paddles
firmly into the paddle wells while discharging.

12. DEFIB MAINT. REQ. message

Contact ZOLL Technical Service Department.

Symptom

Recommended Action

Symptom

Recommended Action

1. The Battery indicator is alternately

illuminating green and yellow.

Verify battery is installed.

Turn unit ON to identify the fault condition.

Replace battery pack with a fully charged battery pack.

If problem persists, replace battery pack, unplug device
from AC mains for more than 10 seconds and plug
device back into AC mains.

2. LOW BATTERY message appears

on monitor when unit is plugged
into AC mains.

Verify that the AC Power Indicator is illuminated.

If not, check AC power cord connection at the wall outlet
and at the rear of the device.

Replace battery pack with a fully charged battery pack.

Unplug device from AC mains, and plug device back
into AC mains.

Verify AC mains is working properly.

3. Neither the Battery, nor AC Power

indicator is illuminated when the
device is plugged into AC mains.

Unplug device from AC mains, and plug device back
into AC mains.

Verify AC mains is working properly.

CHECK CPR PUCK message is
displayed.

No action is required. This is an informational message
to indicate that default calibration values are being used
for the CPR puck. The CPR values will be within the
specification.