By default, the FULLY RELEASE text prompt is not enabled.

This feature is unavailable while using Pediatric CPR electrodes.

CPR Voice Prompts (Adult only)

The R Series unit can be configured to issue voice prompts related to the depth of chest
compressions as feedback to rescuers performing CPR. Two voice prompts are available for
this purpose:

Push Harder

Good Compressions

When chest compressions are detected but their depth is consistently less than 1.5 or 2 inches
(3.8 or 5 cm), depending on the configuration, the defibrillator will issue the prompt “Push
Harder” every 15 seconds. If the rescuer responds by increasing compression depth to more
than 1.5 or 2 inches (3.8 or 5 cm), depending on the configuration, on a consistent basis, the
unit will issue a “Good Compressions” prompt.

See the R Series Configuration Guide for information on enabling/disabling CPR voice
prompts.

CPR Voice prompts are unavailable while using Pediatric CPR electrodes.

Chest Compressions Bar Graph

The R Series unit can display a CPR compression bar graph computed from the CPR sensor
signals. This bar graph, representing depth of compression, is presented on a displacement scale
with a reference marker at 1.5 or 2.0 inches, depending on the configuration. When the full
width of the trace is visible, the unit displays a minimum of 12 seconds of compression data.

Displaying the CPR Bar Graph

To display the CPR displacement bar graph in the Trace 2 or 3 position:

1. Press the Options softkey, then press Traces.

2.

Press either the Trace 2 or Trace 3 softkey.

3.

Press CPR.

Note:

The CPR softkey appears only when OneStep CPR or OneStep Complete electrodes