Pacer – ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual

Page 136

Advertising
background image

C

HAPTER

13

T

ROUBLESHOOTING

13–4

www.zoll.com

9650-0912-01 Rev. K

Pacer

Symptom

Recommended Action

1. CHECK PADS message

Ensure therapy electrodes are connected to the
OneStep cable.

Ensure electrode gel is not dry. Replace therapy
electrodes if necessary.

Ensure good electrode-to-patient contact.

Check integrity of OneStep cable by plugging into test
connector. CHECK PADS should disappear.

2. No stimulus marker (

) is

present on the ECG trace.

Ensure unit is in PACER mode.

Ensure PACER RATE (ppm) is set greater than patient
heart rate.

3. No ventricular capture beat

appears after stimulus marker on
ECG display.

Check patient’s pulse.

Increase output current.

Ensure therapy electrodes are making good contact
with the patient.

Select different ECG Lead configuration.

Review therapy electrode placement.

4. Patient on “Standby” pacing gets

paced intermittently.

Ensure proper ECG electrode or OneStep Pacing/
Complete electrode connection and placement. If ECG
lead wire comes off, pacer will automatically pace
asynchronously.

Check ECG cable for damage.

Patient R-to-R interval varying. Pace rate close to
patient’s heart rate.

Verify rate is set appropriately.

5. Heart rate display reads 0 with

proper pacing capture displayed
on ECG trace.

Check patient’s pulse.

Select different ECG Lead configuration.

6. Bedside/Central Station/Telemetry

ECG display becomes erratic
when pacing.

None, the patient monitor ECG inputs are overloaded by
pacer signals. ECG can only be monitored by the R Series
or pacing device while pacing.

Advertising