5 determine capture – ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual

Page 85

Advertising
background image

Noninvasive Temporary Pacing

9650-0912-01 Rev. K

ZOLL R Series Operator’s Guide

8–5

5 Determine Capture

It is important to recognize when pacing stimulation has produced a ventricular response
(capture). Determination of capture must be assessed both electrically and mechanically in
order to ensure appropriate circulatory support of the patient.

Electrical capture is determined by the presence of a widened QRS complex, the loss of any
underlying intrinsic rhythm, and the appearance of an extended, and sometimes enlarged,
T-wave.

Ventricular response is normally characterized by suppression of the intrinsic QRS complex.

WARNING!

Determination of electrical capture should only be performed by viewing the ECG trace
on the R Series display with its ECG connection directly attached to the patient. Use of
other ECG monitoring devices might provide misleading information due to the
presence of pacer artifacts.

Mechanical capture is assessed by palpation of the peripheral pulse.

To avoid mistaking muscular response to pacing stimuli for arterial pulsations, use ONLY the
following locations for palpating pulse during pacing:

femoral artery

right brachial or radial artery

Effective pacing

The following ECG traces illustrate typical examples of effective pacing.

Changing ECG leads and size can sometimes be helpful in determining capture.

Note:

The shape and size of the paced ECG waveforms can vary depending on the ECG lead
configuration chosen; variation from patient to patient can be expected.

Large T wave

Negative R wave

Inverted T wave/ absence of P waves

Pacer Markers

Pacer Markers

Widened, positive QRS

(which looks like an

ectopic beat).

,

Advertising